The implementation of a manual pretreatment phase increases the efficacy of automated reprocessing procedures in postprocedural bronchoscopes
DOI:
https://doi.org/10.62438/tunismed.v103i5.5237Keywords:
Endoscope reprocessing, , pretreatment phase, automated disinfection, microbial contamination, infection controlAbstract
Introduction: The reprocessing of flexible endoscopes has become a major research topic in
recent decades. The risk of endoscopy-related bacterial, fungal or viral outbreaks is increasing,
particularly those involving multi-resistant strains.
Aim: The aim of the present study was to assess the importance of implementing a manual pre-
treatment phase in conjunction with HDL (High Level Disinfection) cleaning using a Steelco
automated washer-disinfector. The aim of the study was to determine whether neglecting the
manual pre-treatment phase, despite the use of a high-quality automated disinfector, leads to the spread of biofilms and of germs likely to cause healthcare-associated infections.
Methods: To achieve this objective, an extensive microbiological sampling process was carried
out on bronchoscopes under normal cleaning and disinfection conditions prior to the introduction
of the pre-treatment phase.
Based on the results of these samplings, the Quality Control Circle (QCC) team proceeded to
identify and analyze potential causes using the ISHIKAWA diagram during brainstorming
sessions, following which microbiological sampling of the bronchoscopes after implementation
of the results of the cause analysis was carried out.
Results: The results indicate complete elimination of bacterial strains after implementation of
manual pretreatment followed by automated treatment, in line with international disinfection
standards.
Conclusion: The implementation of a pre-treatment phase followed by a Steelco
automated washer-disinfector significantly reduced microbial contamination in bronchoscopes.
Downloads
References
(1) Kovaleva, J., Peters, F. T., van der Mei, H. C., & Degener, J. E. (2013). Transmission of
infection by flexible gastrointestinal endoscopy and bronchoscopy. Clinical microbiology
reviews, 26(2), 231-254. https://doi.org/10.1128%2FCMR.00085-12
(2) McCafferty CE, Aghajani MJ, Abi-Hanna D, Gosbell IB, Jensen SO (2018) An update on
gastrointestinal endoscopy-associated infections and their contributing factors. Ann Clin
Microbiol Antimicrob 17:1-6. https://doi.org/10.1186/s12941-018-0289-2
(3) Kakoullis L, Economidou S, Mehrotra P, Panos G, Karampitsakos T, Stratakos G (2023)
Bronchoscopy-related outbreaks and pseudo-outbreaks: A systematic review. Infect Control
Hosp Epidemiol 1-11. https://doi.org/10.1017/ice.2023.250
(4) Nelson DB, Muscarella LF (2006) Current issues in endoscope reprocessing and infection
control during gastrointestinal endoscopy. World J Gastroenterol 12(25):3953.
https://doi.org/10.3748/wjg.v12.i25.3953
(5) Chang, W. K., Peng, C. L., Chen, Y. W., Sun, C. K., Chen, C. C., Liu, T. C., ... & Chiu, C.
T. (2023). Recommendations and guidelines for endoscope reprocessing: Current position
statement of digestive endoscopic society of Taiwan. Journal of Microbiology, Immunology
and Infection. https://doi.org/10.1016/j.jmii.2023.12.001
(6) Oh, H. J., & Kim, J. S. (2015). Clinical practice guidelines for endoscope reprocessing.
Clinical endoscopy, 48(5), 364-368. http://dx.doi.org/10.5946/ce.2015.48.5.364.
(7) ISO. (2021). ISO 15883-5:2021, Washer-disinfectors — Part 5: Performance requirements
and test method criteria for demonstrating cleaning efficacy
https://www.iso.org/obp/ui/#iso:std:iso:15883:-5:ed-1:v1:en
(8) SF2H (2022) Guide de bonnes pratiques de traitement des dispositifs médicaux réutilisables.
Hygiènes. 2022 ; XXX (5).
(9) Day LW, Kwok K, Visrodia K, Petersen BT (2021) American Society for Gastrointestinal
Endoscopy Infection Control Summit: updates, challenges, and the future of infection
control in GI endoscopy. Gastrointest Endosc 93(1):1-10.
https://doi.org/10.1016/j.gie.2020.06.024
(10) Ministère des affaires sociales et de la sante. (2013). Guide technique traitement des
endoscopes souples thermosensibles à canaux.
sante.gouv.fr/IMG/pdf/dgos_traitement_endoscopes.pdf
(11) Bharti B, Li H, Ren Z, Zhu R, Zhu Z (2022) Recent advances in sterilization and
disinfection technology: A review. Chemosphere 308:136404.
https://doi.org/10.1016/j.chemosphere.2022.136404
(12) Wilson GM (2019) Environmental and pharmaceutical risk factors for the transmission of
Clostridium difficile and other multidrug resistant hospital acquired infections. Dissertation,
University of Iowa. https://doi.org/10.17077/etd.cp4w-j0zs
(13) Maillard JY, Centeleghe I (2023) How biofilm changes our understanding of cleaning and
disinfection. Antimicrob Resist Infect Control 12(1):95. https://doi.org/10.1186/s13756-023-
-4
(14) Cottarelli A, De Giusti M, Solimini AG, Venuto G, Palazzo C, Del Cimmuto A, et al.
(2020) Microbiological surveillance of endoscopes and implications for current reprocessing
procedures adopted by an Italian teaching hospital. Ann Ig 32(2).
https://doi.org/10.7416/ai.2020.2340
(15) Bédard E, Prévost M, Déziel E (2016) Pseudomonas aeruginosa in premise plumbing of
large buildings. Microbiol Open 5(6) : 937-956. https://doi.org/10.1002/mbo3.391
(16) Ofstead, C. L., Wetzler, H. P., Snyder, A. K., & Horton, R. A. (2010). Endoscope
reprocessing methods: a prospective study on the impact of human factors and automation.
Gastroenterology Nursing, 33(4), 304-311.
(17) Forte L, Shum C (2011) Comparative cost-efficiency of the EVOTECH endoscope
cleaner and reprocessor versus manual cleaning plus automated endoscope reprocessing in a
real-world Canadian hospital endoscopy setting. BMC Gastroenterol 11:105.
https://doi.org/10.1186/1471-230x-11-105
(18) Hervé RC, Kong MG, Bhatt S, Chen HL, Comoy EE, Deslys JP, et al. (2023) Evaluation
of cold atmospheric plasma for the decontamination of flexible endoscopes. J Hosp Infect
Additional Files
Published
License
Copyright (c) 2025 La Tunisie Médicale

This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.